Remote control commands
R&S
®
RTO6
1836User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:CAN:SYMBolic:DMIN <SearchName>,<DataPattern>
SEARch:TRIGger:CAN:
SYMBolic:
DMIN? <SearchName>
Defines the minimum data pattern in a symbolic search.
To set the condition, use SEARch:TRIGger:CAN:DCONdition.
Parameters:
<DataPattern> Range: -100E+24 to 100E+24
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName> String that contains the search definition name
SEARch:TRIGger:CAN:SYMBolic:DMAX <SearchName>,<DataPatternTo>
SEARch:TRIGger:CAN:
SYMBolic:DMAX? <SearchName>
Defines the maximum data pattern of the signal in a symbolic search.
This value is required to specify a range if condition INRange | OORange is set with
SEARch:TRIGger:CAN:DCONdition on page 1820.
Parameters:
<DataPatternTo> Range: -100E+24 to 100E+24
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName> String that contains the search definition name
SEARch:TRIGger:CAN:SYMBolic:SGEValue <SearchName>, <SignalEnumValue>
SEARch:TRIGger:CAN:
SYMBolic:SGEValue? <SearchName>
Sets a symbolic data value for signals with enumerated values.
Parameters:
<SignalEnumValue> Numeric value according to the value definition in the DBC file
Parameters for setting and query:
<SearchName> String that contains the search definition name
Example: Definition line in DBC file:
VAL_ 2175091489 Gear 0 "Idle" 1 "Gear_1" 2 "Gear_2"
3 "Gear_3" 4 "Gear_4" 5 "Gear_5" ;
Search for "Gear_3"
SEARch:TRIGger:CAN:SYMBolic:SGEValue "Search1",3