Remote control commands
R&S
®
RTO6
1531User Manual 1801.6687.02 ─ 05
<n> 1..18
Number of the histogram measurement, see T
able 24-10.
Parameters:
<UpperLimit>
Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
MEASurement<m>:HISTogram:LCHeck<n>:LOWer:MARGin <LowerMargin>
MEASurement<m>:
HIST
ogram:LCHeck<n>:UPPer:MARGin <UpperMargin>
Define the lower and upper margins for the limit check, respectively. Margins are not as
strict as limits and must be within the valid value range. The valid range is defined
using the MEASurement<m>:HISTogram:LCHeck<n>:VALid command.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..18
Number of the histogram measurement, see Table 24-10.
Parameters:
<UpperMargin>
Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
24.12.9 Display
MEASurement<m>:DISPlay:LEV
els............................................................................... 1531
MEASurement<m>:DISPlay:RESults..............................................................................1532
MEASurement<m>:DISPlay:STYLe............................................................................... 1532
MEASurement<m>:DISPlay:HISTogram.........................................................................1532
MEASurement<m>:DISPlay:LEVels <DisplayLevels>
If enabled, the reference levels used for the measurement are displayed in the dia-
gram.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page
1479.
Parameters:
<DisplayLevels>
ON | OFF
*RST: OFF
Automatic measurements