Loading...
Remote control commands
R&S
®
RTO6
1344User Manual 1801.6687.02 ─ 05
Predefined probes...............................................................................................
1361
Current probes.................................................................................................... 1362
Probe attributes
...................................................................................................1363
Probe adapter R&S RTZ2T................................................................................1365
24.8.7.1 Common probe settings
PROBe<m>:SETup:STATe?.......................................................................................... 1344
PROBe<m>:SETup:TYPE?...........................................................................................1344
PROBe<m>:SETup:NAME?.......................................................................................... 1344
PROBe<m>:SETup:BANDwidth?...................................................................................1345
PROBe<m>:SETup:ATTenuation[:AUTO]?......................................................................1345
PROBe<m>:SETup:OFFSet:AZERo...............................................................................1345
PROBe<m>:SETup:OFFSet:USEautozero......................................................................1346
CHANnel<m>:EATScale............................................................................................... 1346
CHANnel<m>:EATTenuation......................................................................................... 1346
PROBe<m>:SETup:STATe?
Queries if the probe at the specified input channel is active (detected) or not active (not
detected). To switch the probe on, use CHANnel<m>:STATe.
Suffix:
<m>
.
1..4
Return values:
<State> DETected | NDETected
*RST: NDETected
Usage: Query only
Asynchronous command
PROBe<m>:SETup:TYPE?
Queries the type of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Type> String containing one of the following values:
None (no probe detected)
Passive Probe
active single-ended
Usage: Query only
Asynchronous command
PROBe<m>:SETup:NAME?
Queries the name of the probe.
Acquisition and setup