Remote control commands
R&S
®
RTO6
1365User Manual 1801.6687.02 ─ 05
Return values:
<InputImpedance> Range: 0 to 1E+9
*RST: 50
Default unit: Ω
Usage: Query only
Asynchronous command
24.8.7.10 Probe adapter R&S RTZ2T
PROBe<m>:SETup:ADAPter?
.......................................................................................1365
PROBe<m>:SETup:ATTenuation:TDEFprobe..................................................................1365
PROBe<m>:SETup:ADAPter?
Queries the adapter status, whether the instrument identified the adapter.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Type> NONE | Z2T
*RST: NONE
Usage: Query only
Asynchronous command
PROBe<m>:SETup:ATTenuation:TDEFprobe <TekPredefProbe>
Selects the T
ektronix probe that is connected to the R&S RTZ2T adapter.
Suffix:
<m>
.
1..4
Parameters:
<TekPredefProbe> NONE | P5205A50 | P5205A500 | P5210A100 | P5210A1000 |
P6205 | P6241 | P6243 | P6245 | P6246A1 | P6246A10 |
P6247A1 | P6247A10 | P6248A1 | P6248A10 | P6249 |
P6250A5 | P6250A50 |
P6251A5 | P6251A50 | P6701B |
P6703B | P6711 | P6713 | TCP202
*RST: NONE
Example:
PROBe2:SETup:ADAPter
<-- Z2T
PROBe2:SETup:ATTenuation:TDEFprobe P5205A50
Checks the adapter state on channel 2, and selects the probe
P5205A50.
Usage: Asynchronous command