Remote control commands
R&S
®
RTO6
1733User Manual 1801.6687.02 ─ 05
24.17.5.2 T
rigger
The trigger suf
fix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:UART:TYPE.............................................................................................1733
TRIGger<m>:UART:SOURce........................................................................................ 1733
TRIGger<m>:UART:DPOPerator....................................................................................1733
TRIGger<m>:UART:DPOSition...................................................................................... 1734
TRIGger<m>:UART:DPTO............................................................................................ 1734
TRIGger<m>:UART:FCONdition.................................................................................... 1734
TRIGger<m>:UART:DATA.............................................................................................1734
TRIGger<m>:UART:TYPE <Type>
Selects the trigger type for UART analysis.
See also: "Type" on page 525
Parameters:
<Type> STBT | PCKS | DATA | PRER | BRKC | STPerror
STBT
: Start bit
PCKS: Packet start
DA
TA: Serial pattern
PRER: Parity error
BRKC: Break condition
STPerror: Stop error
*RST: STBT
Usage: Asynchronous command
TRIGger<m>:UART:SOURce <Source>
Selects the transmitter or receiver line as trigger source.
Parameters:
<Source> TX | RX
*RST: TX
Usage: Asynchronous command
TRIGger<m>:UART:DPOPerator <DataPosOperator>
Sets the operator for the data position. You can defined an exact position, or a position
range.