Remote control commands
R&S
®
RTO6
1732User Manual 1801.6687.02 ─ 05
Firmware/software: FW 2.25
BUS<m>:UART:TOUT <InterframeT
ime>
Sets the timeout between packets in a UAR
T data stream. A new packet starts with the
first start bit after the timeout.
The command is relevant if BUS<m>:UART:PACKets is set to TOUT.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<InterframeTime> Range: 1E-6 to 1
Increment: 1
*RST: 1E-3
Default unit: s
Usage: Asynchronous command
Firmware/software: FW 2.25
BUS<m>:UART:EWORd <EndOfFrame>
Sets the end pattern of the packets. A new packet starts with the first start bit after the
defined end pattern.
The command is relevant if BUS<m>:UART:PACKets
is set to EWORd.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<EndOfFrame> End word value in decimal format (range 0 to 255) or hexadeci-
mal format (prefix #H). The query always returns hexadecimal
values.
Example:
:BUS:UART:PACK EWOR
:BUS:UART:
EWOR 10 // Decimal value
:BUS:UART:EWOR?
#H0A // Query returns hex
Example:
:BUS:UART:PACK EWOR
:BUS:UART:EWOR #Hff
// Hexadecimal, prefix #H
:BUS:UART:EWOR?
#HFF
Usage: Asynchronous command