Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1156User Manual 1801.6687.02 ─ 05
19.4.2.1 Dynamic on resistance results
The results of "Dynamic on Resistance"
measurements are provided in two ways:
●
The diagram shows the graphical presentation of:
– the voltage waveform
– the current waveform
●
The result box displays the dynamic on resistance value.
To measure and display the dynamic on resistance, the instrument uses the following
measurements and waveforms:
●
"MG 5" to measure the amplitude of the voltage
●
"MG 6"
to measure the amplitude of the voltage
●
"MG 7" to measure the amplitude of the current
●
"MG 8" to measure the amplitude of the current
●
"Cu 1 " Cursor 1 to measure gate ["t
0
" ,"t
1
"]
●
"Cu 2 " Cursor 2 to measure gate ["t
2
" ,"t
3
"]
The used resources are listed in the "Details" tab. See also: Chapter 19.2.3, "Details
tab", on page 1135.
The dynamic on resistance displayed as the result is defined as:
The points "t
0
" , "t
1
", "t
2
" and "t
3
" are defined by the cursor lines displayed in the result
diagram of the measurement. Y
ou can move the cursor lines to define another area of
interest.
The following remote commands are used for handling the measurement results:
●
POWer:
DONRes:RESult:RESistance? on page 2536
●
POWer:DONRes:GATE<m>:STARt on page 2536
●
POWer:DONRes:GATE<m>:STOP on page 2536
●
POWer:DONRes:REPort:ADD on page 2536
19.4.2.2 Configuring dynamic on resistance
For details of the configuration settings, see Chapter 19.4.2.3, "Dynamic on resistance
settings", on page 1158.
Control loop