14 Mixed signal option (MSO, R&S RTO6-B1)
The Mixed Signal Option R&S RT
O6-B1 adds logic analyzer functions to the classical
oscilloscope functions. Using the MSO option, you can analyze and debug embedded
systems with mixed-signal designs that use analog signals and correlated digital sig-
nals simultaneously.
The Mixed Signal Option provides 16 digital channels grouped in two logic probes
(pods) with 8 channels each. The instrument ensures that analog and digital wave-
forms are time-aligned and synchronized so that critical timing interactions between
analog and digital signals can be displayed and tested. The automatic alignment com-
pensates the skew between the probe connectors of the analog channels and the
probe boxes of the digital channels.
Logic probes provided by Rohde & Schwarz are recognized by the instrument. The
information on each recognized probe is shown under "Menu"
"Settings" > "Mainte-
nance" > "Hardware" > "Advanced" > "Logic Probes".