24.8.7.2 Micro button and R&S ProbeMeter
PROBe<m>:SET
up:MODE
........................................................................................... 1347
PROBe<m>:SETup:DISPlaydiff..................................................................................... 1348
PROBe<m>:PMETer:VISibility.......................................................................................1348
PROBe<m>:PMETer:RESults:SINGle?...........................................................................1349
PROBe<m>:PMETer:RESults:COMMon?....................................................................... 1349
PROBe<m>:PMETer:RESults:DIFFerential?................................................................... 1349
PROBe<m>:PMETer:RESults:NEGative?....................................................................... 1350
PROBe<m>:PMETer:RESults:POSitive?........................................................................ 1350
PROBe<m>:SETup:MODE <Mode>
Select the action that is started when you press the micro button on the probe head.
See also: "Micro button action" on page 148.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Mode> RCONtinuous | RSINgle | AUToset | AZERo | SEToffsettomean |
SITFile |
NOACtion | FINDtriglevel |
REPort | PROBemode |
PRSetup
RCONtinuous
Run continuous: The acquisition is running as long as the probe
button is pressed.
RSINgle
Run single: Starts a defined number of acquisitions (same as
[Single] key).
AUTOSET
Starts the autoset procedure.
AZero
AutoZero: performs an automatic correction of the zero error.
SEToffsettomean
Set offset to mean: performs an automatic compensation for a
DC component of the input signal.
SITFile
Save Image To File:
Directs the display image to a file. The MMEMory:NAME com-
mand defines the file name. The file format is defined with
HCOPy:DEVice<m>:LANGuage.
NOACtion
Nothing is started on pressing the micro button.
FINDtriglevel
Sets the trigger level automatically to 0.5 * (MaxPeak – Min-
Peak). The function is not available for an external trigger
source.