Remote control commands
R&S
®
RTO6
1349User Manual 1801.6687.02 ─ 05
PROBe<m>:PMETer:RESults:SINGle?
Returns the ProbeMeter measurement result of single-ended active R&S probes, the
voltage measured between the probe tip and the ground.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Result> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Firmware/software: FW 2.25
PROBe<m>:PMETer:RESults:COMMon?
Returns the ProbeMeter measurement result of differential active R&S probes: the
common mode voltage, which is the mean voltage between the signal sockets and the
ground socket.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Result> Range:
-100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Firmware/software: FW 2.25
PROBe<m>:PMETer:RESults:DIFFerential?
Returns the ProbeMeter measurement result of differential active R&S probes, the dif-
ferential voltage - the voltage between the positive and negative signal sockets.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Dif
fMeasRes> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Acquisition and setup