Remote control commands
R&S
®
RTO6
1352User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<GainManual> Range: 100E-6 to 10000
Increment: 100E-6
*RST: 1
Default unit: V/A
Usage: Asynchronous command
24.8.7.4 Active voltage probes
PROBe<m>:SETup:CMOFfset
...................................................................................... 1352
PROBe<m>:SETup:ZAXV.............................................................................................1352
PROBe<m>:SETup:ACCoupling....................................................................................1353
PROBe<m>:SETup:ADVanced:PMToffset.......................................................................1353
PROBe<m>:SETup:ADVanced:RANGe..........................................................................1353
PROBe<m>:SETup:ADVanced:FILTer............................................................................ 1354
PROBe<m>:SETup:ADVanced:AUDioverload................................................................. 1354
PROBe<m>:SETup:CMOFfset <CMOffset>
Sets the common-mode offset. The setting is only available for differential probes.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<CMOffset> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:ZAXV <ExtAttRTZA15>
If you use the external attenuator R&S R
T-ZA15 together with one of the differential
active probes R&S RT-ZD10/20/30, enable it to include the external attenuation in the
measurements.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAttRTZA15> ON | OFF
*RST: OFF
Acquisition and setup