Remote control commands
R&S
®
RTO6
1353User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
PROBe<m>:SETup:ACCoupling <ProbeCouplingAC>
Enables AC coupling in the R&S R
T
‑ZPR20 probe, which removes DC and very low-
frequency components. The R&S RT‑ZPR20 probe requires 50 Ω input termination,
for which the channel AC coupling is not available. The probe setting allows AC cou-
pling also at 50 Ω inputs.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeCouplingAC> ON | OFF
*RST: OFF
Usage: Asynchronous command
PROBe<m>:SETup:ADVanced:PMToffset
Sets the measured ProbeMeter value as offset. Make sure that the ProbeMeter is
active before you use this command.
Only available for power rail probes R&S
RT‑ZPR.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
PROBe<m>:SETup:ADVanced:RANGe <ProbeRange>
Sets the voltage range of a R&S RT
-ZHD probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeRange> AUTO | MHIGh | MLOW
AUTO
The voltage range is set with CHANnel<m>:SCALe.
MHIGh
Sets the higher voltage range of the connected probe. To query
the value, use PROBe<m>:SETup:ATTenuation[:AUTO]?.
MLOW
Sets the lower voltage range of the connected probe. To query
the value, use PROBe<m>:SETup:ATTenuation[:AUTO]?.
Acquisition and setup