EasyManua.ls Logo

R&S RTO6 - 24.18 Mixed signal option (MSO, R&S RTO6-B1); 24.18.1 Digital channels

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2420User Manual 1801.6687.02 ─ 05
Parameters:
<ZoomCoupling> ON | OFF
*RST: OFF
Usage: Asynchronous command
EYE<m>:ZDIagram <ZoomDiagram>
Selects the zoom to which the eye mask area is set if EYE<m>:ZCOupling is enabled.
Suffix:
<m>
.
1..4
Parameters:
<ZoomDiagram> String with the name of the diagram on which the zoom is based
Usage: Asynchronous command
24.18 Mixed signal option (MSO, R&S RT
O6-B1)
This chaptes decribes the remote commands of MSO option R&S
RTO6-B1.
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
Chapter 24.3, "Programming examples", on page 1259
Digital channels...................................................................................................2420
Parallel bus configuration....................................................................................2423
Digital resolution..................................................................................................2430
Trigger settings for digital signals and parallel buses......................................... 2431
MSO data............................................................................................................2440
24.18.1 Digital channels
All DIGital: commands affect only the settings of the first MSO bus (Bus1). The set-
tings of all other parallel buses (Bus 2, 3,4) remain unchanged.
DIGital<m>:DISPlay..................................................................................................... 2421
DIGital<m>:TECHnology...............................................................................................2421
DIGital<m>:THReshold.................................................................................................2421
DIGital<m>:THCoupling................................................................................................2422
Mixed signal option (MSO, R&S R
T
O6-B1)

Table of Contents

Other manuals for R&S RTO6