EasyManua.ls Logo

R&S RTO6 - 24.16.5 Waveform histogram export to file

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1656User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
EXPort:WAVeform:DISPlayoff <FastExport>
Enables or disables the display update during an Nx Single acquisition.
Parameters:
<FastExport> ON | OFF
ON: Disables the display update for maximum export speed.
OFF: Enables the display update. The export is slower.
*RST
: OFF
Usage: Asynchronous command
EXPort:WAVeform:FASTexport <Enable>
To improve the performance of data export to file, the measurements are performed
slower while the data export speeds up.
Setting parameters:
<Enable> ON |
OFF
Example: See Chapter 24.3.5.2, "Exporting waveform data to file",
on page 1267
Usage: Setting only
Asynchronous command
24.16.5 Waveform histogram export to file
EXPort:HIST
ogram:SELect............................................................................................1656
EXPort:HISTogram:INCidence.......................................................................................1656
EXPort:HISTogram:NAME.............................................................................................1657
EXPort:HISTogram:SAVE..............................................................................................1657
EXPort:HISTogram:DATA?............................................................................................1657
EXPort:HISTogram:SELect <Name>
Selects the histogram to be exported.
Parameters:
<Name> String with the histogram name.
Example: See "Exporting histogram data to file" on page
1264
Usage: Asynchronous command
EXPort:HISTogram:INCidence <Incidence>
Sets the mode of exported data: relative or absolute frequency of amplitude values.