Waveform analysis
R&S
®
RTO6
257User Manual 1801.6687.02 ─ 05
The vertical scale of a math waveform is adapted automatically to the measurement
results to ensure optimal display
. Furthermore, you can scale each math waveform
manually in vertical direction like a channel waveform.
As for channel waveforms, you can also change the arithmetic mode for the waveform
to display the envelope or an average over several calculations.
Y
ou can store a math waveform as a reference waveform and restore it later, see "To
save a reference waveform"
on page 250.
Displaying math waveforms.................................................................................. 257
Math setup - general settings................................................................................258
Operator editor......................................................................................................259
Equation................................................................................................................261
Filters.................................................................................................................... 269
Math scale settings............................................................................................... 271
Math arithmetic......................................................................................................272
7.3.1 Displaying math waveforms
Math waveforms can be displayed in addition to the channel and other waveforms.
They also can be used for analysis, e.g. measurements, even if the math waveform is
not active.
1.
In the "Math" menu, select "Math Setup".
Alternatively
, press the [Math] key.
2. Define the math expression for calculation in one of the following ways:
Chapter 7.3.3.2, "Defining a formula in the Operator editor", on page 260
Chapter 7.3.4, "Equation", on page 261
Chapter 9.1.2, "Configuring spectrum waveforms", on page 374
3. In the "Math Setup" dialog box, in the "Setup" tab, tap the "Enable math signal".
The math waveform is displayed on the screen.
4. To change the vertical scaling of the math waveform, tap the "Manual" icon.
5. Enter the "Vertical scale" factor (per division). If necessary, add a "Vertical offset".
By default the instrument performs an automatic scaling.
Tip: You can also use the vertical [Scale] rotary knob for scaling. In this case, the
scale mode is set to "Manual" temporarily.
6. If you need the envelope or average of the math waveform over several calcula-
tions, change the arithmetic mode for the waveform as for channel waveforms.
See also: "Arithmetic" on page 130.
7. Close the "Math Setup" dialog box.
Mathematics