Remote control commands
R&S
®
RTO6
1361User Manual 1801.6687.02 ─ 05
PROBe<m>:DEEMbedding:LOAD:IMPedance <SourceImpedance>
Shows the source impedance of the DUT
. If an R&S
RT‑ZM probe is connected, the
value depends on the selected probe mode: common mode, differential, or single-
ended measurement.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<SourceImpedance> Range: 100E-15 to 1E+9
Increment: 1
*RST: 50
Default unit: Ω
Usage: Asynchronous command
24.8.7.7 Predefined probes
PROBe<m>:SETup:A
TTenuation:DEFProbe................................................................... 1361
PROBe<m>:SETup:OFFSet:TOMean............................................................................ 1362
PROBe<m>:SETup:ATTenuation:DEFProbe <PredefinedProbe>
Selects a predefined probe. These are probes that are not recognized automatically
but the parameters of the probe are known to the instrument.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<PredefinedProbe> ZC10 | ZC20 | ZC30 | ZD01A100 | ZD01A1000 | ZZ80 | FREE |
ZS10L | ZD02 |
ZD08 | ZC02100 | ZC021000 |
ZC03 |
ZD002A10 | ZD002A100 | ZD003A20 | ZD003A200 | ZC3110 |
ZC311 | ZC3101
ZC10 | ZC20 | ZC30 | ZZ80 | ZS10L | ZD02 | ZD08 | ZC03
Type of the probe
ZD01A100 | ZD01A1000 | ZD002A10 | ZD002A100 |
ZD003A20 | ZD003A200
High voltage differential probes R&S RT-ZD0xx, attenuation ratio
according to the setting on the probe.
A10 = 10:1
A20 = 20:1
A100 = 100:1
A200 = 200:1
A1000 = 1000:1
ZC02100 | ZC021000
Current probes 100 A or 1000 A according to the setting on the
probe.
Acquisition and setup