Remote control commands
R&S
®
RTO6
1396User Manual 1801.6687.02 ─ 05
24.9.5 Runt trigger
TRIGger<m>:
RUNT
:POLarity........................................................................................ 1396
TRIGger<m>:LEVel<n>:RUNT:LOWer............................................................................1396
TRIGger<m>:LEVel<n>:RUNT:UPPer............................................................................ 1396
TRIGger<m>:RUNT:RANGe..........................................................................................1396
TRIGger<m>:RUNT:WIDTh...........................................................................................1397
TRIGger<m>:RUNT:DELTa........................................................................................... 1397
TRIGger<m>:RUNT:POLarity <Polarity>
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Polarity> POSitive | NEGative | EITHer
See Chapter 24.4.4, "Polarity parameter", on page 1285.
*RST
: POSitive
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:RUNT:LOWer <Level>
TRIGger<m>:LEVel<n>:
RUNT:UPPer <Level>
Set the lower and upper voltage thresholds.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
<n> 1..31
Indicates the trigger source:
1...4 = channel 1...4
5...27 = not available
28...31 = differential signals Dif
f1, Diff2 and Common1, Com-
mon2
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: Lower = -0.1, upper =0.1
Default unit: V
Usage: Asynchronous command
TRIGger<m>:RUNT:RANGe <Mode>
Defines the time limit of the runt pulse in relation to the TRIGger<m>:RUNT:
WIDTh
and TRIGger<m>:RUNT:DELTa settings.
Trigger