Remote control commands
R&S
®
RTO6
1395User Manual 1801.6687.02 ─ 05
WITHin
T
riggers on pulses inside a given range. The range is defined by
the width 
±delta.
OUTSide
Triggers on pulses outside a given range. The range is defined
by the width ±delta.
SHORter
Triggers on pulses shorter than the given width.
LONGer
Triggers on pulses longer than the given width.
*RST:  WITHin
Usage:  Asynchronous command
TRIGger<m>:WIDTh:WIDTh <Width>
For the ranges "Within" and "Outside" (defined using TRIGger<m>:WIDTh:RANGe),
the width defines the center of a range which is defined by the limits "±Delta" (see
TRIGger<m>:WIDTh:DELTa on page 1395).
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
pulse width, respectively.
Suffix: 
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Width> Range:  100E-12  to  10000
Increment:  100E-9
*RST:  5E-9
Default unit: s
Usage:  Asynchronous command
TRIGger<m>:WIDTh:DELTa <WidthDelta>
Defines a range around the width value specified using TRIGger<m>:WIDTh:
WIDTh.
Suffix: 
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<WidthDelta> Range:  0  to  432
Increment:  500E-12
*RST:  0
Default unit: s
Usage:  Asynchronous command
Trigger