Remote control commands
R&S
®
RTO6
1970User Manual 1801.6687.02 ─ 05
Parameters:
<TypeSpecifier> CSTatus | DATA | ALL
CST
atus: command/status word
*RST
: ALL
24.17.13.3 Decode results
To load and acivate a label list, use:
●
BUS<m>:
NEWList on page 1672
●
BUS<m>:SYMBols on page 1672
BUS<m>:MILStd:WCOunt?...........................................................................................1970
BUS<m>:MILStd:WORD<n>:DATA?.............................................................................. 1970
BUS<m>:MILStd:WORD<n>:INFO?...............................................................................1971
BUS<m>:MILStd:WORD<n>:RTADdress?...................................................................... 1971
BUS<m>:MILStd:WORD<n>:STARt?............................................................................. 1971
BUS<m>:MILStd:WORD<n>:STATus?........................................................................... 1972
BUS<m>:MILStd:WORD<n>:STOP?..............................................................................1972
BUS<m>:MILStd:WORD<n>:SYMBol?...........................................................................1972
BUS<m>:MILStd:WORD<n>:TYPE?..............................................................................1973
BUS<m>:MILStd:WCOunt?
Returns the number of decoded words.
Suffix:
<m>
.
1..4
Return values:
<FrameCount> Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:MILStd:WORD<n>:DATA?
Return the data bytes of the specified word.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the word.
Return values:
<Data> 16-bit data of the specified word as a 2-byte bit pattern (B1, B2).
The first byte B1 is the most significant byte.
Example:
BUS:MILStd:WORD4:DATA?
<-- #H08,
#H49
Protocols