13.8.2.3 Configuring LIN signals
For configuration, assign the lines to the input channels, and define the active states
and the logical thresholds.
For details on configuration settings, see Chapter 13.8.2.1, "LIN configuration settings",
on page
606.
1.
Press the [Bus] key on the front panel.
2. At the left hand-side, select the vertical tab of the bus you want to set up.
3. Select the tab of the bus you want to set up, for example "SB1".
4. Tap the "Protocol" button and select the protocol: "LIN".
5. Set the "Data source".
6. Set the logical thresholds.
7. Set the "LIN standard", "Bit rate" and "Polarity" according to your signal.
8. Enable "Decode".
13.8.3 LIN trigger
13.8.3.1 LIN trigger settings
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
LIN (option R&S RT
O6-K520)