Protocol analysis
R&S
®
RTO6
611User Manual 1801.6687.02 ─ 05
Make sure that:
●
In the [Bus]
> "Setup" tab:
–
The correct "Protocol" is selected.
– The data sources of the serial bus are channel signals.
●
In the "Trigger" > "Setup" tab:
– The "Trigger on" = "Single event"
– The trigger "Source" = "Serial bus".
– The correct "Serial bus" is selected.
Trigger type
Selects the trigger type for LIN analysis.
"Start of frame
(Sync)"
Triggers on the stop bit of the sync field.
"Identifier"
Sets the trigger to one specific identifier or an identifier range. Enter
only the 6-bit identifier without parity bits, not the protected identifier.
Description of trigger type specific settings: "Identifier setup: Condi-
tion, Frame ID min, Frame ID max"
on page 612.
"Identifier OR"
Sets the trigger to a combination of up to four identifiers.
Description of trigger type specific settings: "Identifier OR setup: Mon-
itor, Frame ID" on page 613
LIN (option R&S RT
O6-K520)