Protocol analysis
R&S
®
RTO6
572User Manual 1801.6687.02 ─ 05
13.6.3.2 T
riggering on custom Manchester / NRZ serial bus
Prerequisite: A bus is configured for the custom serial bus signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [Bus]
key and select the "Trigger" tab.
2. Press the "Select" button. The "Source" button is shown.
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to Custom.
5. Tap "Type" and select the trigger type to be used for custom serial bus protocol
analysis.
6. Depending on the selected custom serial bus coding standard, more setup condi-
tions have to be specified.
For information on how to proceed with the configuration settings, see Chap-
ter 13.6.3.1, "Custom: Manchester / NRZ trigger settings", on page 567.
13.6.4 Custom filter
With the filter function you can display only selected events in the acquired data. Only
one single filter can be enabled on the protocol at a time.
13.6.4.1 Custom filter settings
Access:
1.
Press [Bus].
2.
Set "Protocol = Custom".
3. In the "Display" tab, enable "Apply filter".
4. Open "Setup filter".
Custom: Manchester / NRZ (option R&S RT
O6-K510)