Remote control commands
R&S
®
RTO6
1594User Manual 1801.6687.02 ─ 05
MTESt:WFMRescale:YPOSition <MaskT
estName>,<V
ertPosi>
MTESt:WFMRescale:YPOSition? <MaskTestName>
Moves the mask vertically within the display.
Parameters:
<VertPosi> Range: -1000 to 1000
Increment: 0.01
*RST: 0
Default unit: div
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
Usage: Asynchronous command
MTESt:WFMRescale:YSTRetch <MaskTestName>,<V
erticalStretch>
MTESt:WFMRescale:YSTRetch? <MaskTestName>
Sets the vertical scaling to stretch the mask in y-direction. The scaling axis is the hori-
zontal line through the lowest value of the lower mask limit.
Parameters:
<VerticalStretch> Scale factor in %
Range: 10 to 1000
Increment: 1
*RST: 100
Default unit: %
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
Usage: Asynchronous command
24.14.4 Event actions
MTESt:ONV
iolation:BEEP.............................................................................................1594
MTESt:ONViolation:STOP.............................................................................................1595
MTESt:ONViolation:PRINt.............................................................................................1595
MTESt:ONViolation:SAVewaveform............................................................................... 1595
MTESt:ONViolation:REPort...........................................................................................1596
MTESt:ONViolation:TRIGgerout.................................................................................... 1596
MTESt:ONViolation:RUNexec....................................................................................... 1596
MTESt:ONViolation:BEEP <MaskTestName>,<Beep>
MTESt:ONViolation:BEEP? <MaskTestName>
Generates a beep sound for the specified event.
Mask testing