EasyManua.ls Logo

R&S RTO6 - 24.17.14.3 Decode results

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1989User Manual 1801.6687.02 ─ 05
Parameters:
<MinGapBits> Range: 0 to 100
Increment: 1
*RST: 4
Default unit: bit
TRIGger<m>:ARINc:MAXGap:SELect <MaxGapSelect>
Enables triggering when the maximum gap is out of range. You can set the maximum
gap with:
TRIGger<m>:ARINc:MAXGap:BITS.
Parameters:
<MaxGapSelect> ON | OFF
*RST: OFF
TRIGger<m>:ARINc:MAXGap:BITS <MaxGapBits>
Sets the value for the maximum gap to be triggered on.
Parameters:
<MaxGapBits> Range: 0 to 1000
Increment: 1
*RST: 100
Default unit: bit
TRIGger<m>:ARINc:SDI <SDI>
Sets the source/destination identifier (SDI) bits.
Parameters:
<SDI>
TRIGger<m>:ARINc:SSM <SSM>
Sets the sign/status matrix (SSM) bits.
Parameters:
<SSM>
24.17.14.3 Decode results
To load and acivate a label list, use:
BUS<m>:
NEWList on page 1672
BUS<m>:SYMBols on page 1672
BUS<m>:ARINc:WCOunt?............................................................................................1990
BUS<m>:ARINc:WORD<n>:DATA?............................................................................... 1990
BUS<m>:ARINc:WORD<n>:LABel?...............................................................................1990
BUS<m>:ARINc:WORD<n>:PATTern?........................................................................... 1991