Remote control commands
R&S
®
RTO6
1933User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<ExpectedIFSMax> Range: 16 to 100000
Increment: 1
*RST: 1000
BUS<m>:CXPI:IFS:MIN <ExpectedIFSMin>
Sets the minimum for the inter-frame-space range.
Suffix:
<m>
.
1..4
Parameters:
<ExpectedIFSMin> Range: 16 to 1000
Increment: 1
*RST: 21
24.17.12.2 Trigger
The trigger suf
fix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
TRIGger<m>:SOURce[:SELect] is set to SBUS.
●
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
●
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:CXPI:CT..................................................................................................1934
TRIGger<m>:CXPI:DATA:DCONdition............................................................................1934
TRIGger<m>:CXPI:DATA:DMAX....................................................................................1934
TRIGger<m>:CXPI:DATA:DMIN.....................................................................................1935
TRIGger<m>:CXPI:DATA:ICONdition............................................................................. 1935
TRIGger<m>:CXPI:DATA:IMAX.....................................................................................1935
TRIGger<m>:CXPI:DATA:IMIN...................................................................................... 1935
TRIGger<m>:CXPI:DEXTension:CONDition....................................................................1936
TRIGger<m>:CXPI:DEXTension:MAX............................................................................ 1936
TRIGger<m>:CXPI:DEXTension:MIN............................................................................. 1936
TRIGger<m>:CXPI:DLC:CONDition............................................................................... 1937
TRIGger<m>:CXPI:DLC:MAX........................................................................................1937
TRIGger<m>:CXPI:DLC:MIN.........................................................................................1937
TRIGger<m>:CXPI:ERRor:CRC.................................................................................... 1937
TRIGger<m>:CXPI:ERRor:DLC.....................................................................................1938
TRIGger<m>:CXPI:ERRor:IBS...................................................................................... 1938
TRIGger<m>:CXPI:ERRor:IFS...................................................................................... 1938
TRIGger<m>:CXPI:ERRor:LENGth................................................................................1938
TRIGger<m>:CXPI:ERRor:PARity..................................................................................1939
TRIGger<m>:CXPI:ERRor:UART...................................................................................1939
Protocols