Remote control commands
R&S
®
RTO6
1938User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..3
Parameters:
<ErrorCRC> ON | OFF
*RST: ON
TRIGger<m>:CXPI:ERRor:DLC <ErrorDLC>
Triggers on a data length code (DLC) error
, if TRIGger<m>:CXPI:TYPE is set to
ERRor.
DLC error occurs when the value of the DLC field is different from the data byte field
value.
Suffix:
<m>
.
1..3
Parameters:
<ErrorDLC> ON | OFF
*RST: ON
TRIGger<m>:CXPI:ERRor:IBS <ErrorIBS>
Triggers on an inter-byte-space (IBS) error, if TRIGger<m>:CXPI:TYPE is set to
ERRor.
Suffix:
<m>
.
1..3
Parameters:
<ErrorIBS> ON | OFF
*RST: ON
TRIGger<m>:CXPI:ERRor:IFS <ErrorIFS>
Triggers on an inter-frame-space (IFS) error
, if TRIGger<m>:CXPI:TYPE is set to
ERRor.
Suffix:
<m>
.
1..3
Parameters:
<ErrorIFS> ON | OFF
*RST: ON
TRIGger<m>:CXPI:ERRor:LENGth <ErrorLength>
Triggers on a length error
, if TRIGger<m>:CXPI:TYPE is set to ERRor.