Pulse source and TDR/TDT analysis (options R&S RTO6-B7/K130)
R&S
®
RT
O6
1053User Manual 1801.6687.02 ─ 05
Figure 16-3: Calibration of single-ended TDR/TDT
The used short type is shown in the wizard. To change the short, tap the "Edit" icon.
Short type
Selects whether the short is the Rohde & Schwarz short that is delivered with the pulse
source R&S R
T
O6-B7, or another one.
If you do not use the Rohde & Schwarz short, enter the "Electrical short length" of the
used short.
Remote command:
TDRT:SLENgth on page 2467
Calibrate
Starts the calibration of the current calibration step (short, open, match through).
16.2.4 Calibration results
Access: [App Cockpit]> "Analysis" > "TDR / TDT" > "Calibrate setup" > "Next"