Remote control commands
R&S
®
RTO6
2467User Manual 1801.6687.02 ─ 05
Usage: Event
Asynchronous command
TDRT:TYPE <T
ype>
Selects the analysis type: TDR (time domain reflectometry), TDT (time domain trans-
mission), or TDR/TDT (both TDR and TDT).
Parameters:
<T
ype> TDR | TDRTdt | TDT
*RST: TDR
Usage: Asynchronous command
TDRT:PULSe:OUT <Channel>
Sets the channel to which the "Out"
connector of the pulse source is connected via cal-
ibration standard or DUT.
Parameters:
<Channel> C1W1 | C2W1 | C3W1 | C4W1
*RST: C2W1
Usage: Asynchronous command
TDRT:PULSe:REFerence <Channel>
Sets the channel to which the "Ref" connector of the pulse source is connected.
Parameters:
<Channel> C1W1 | C2W1 |
C3W1 | C4W1
*RST: C1W1
Usage: Asynchronous command
TDRT:SLENgth <ShortLength>
Sets the electrical length of a user-defined short.
The reset value is the electrical length of the short hat is delivered with the pulse
source R&S RT
O6-B7.
Parameters:
<ShortLength> Range: 0 to 2E-9
Increment: 1E-12
*RST: 26.5E-12
Default unit: s
Usage: Asynchronous command
Pulse Source and TDR/TDT Analysis (options R&S RT
O6-B7/K130)