Triggers
R&S
®
RTO6
222User Manual 1801.6687.02 ─ 05
"Events"
Defines the holdoff as a number of trigger events. The next trigger
only occurs when this number of events is reached. The number of
triggers to be skipped is defined in
"Holdof
f events".
"Random"
Defines the holdoff as a random time limited by "Minimum time" and
"Maximum time". For each acquisition cycle, the instrument selects a
new random holdoff time from the specified range.
Random holdof
f prevents synchronization to discover effects invisible
with synchronized triggering, for example, the features of a pulse
train.
"Auto"
The holdoff time is calculated automatically based on the current hori-
zontal scale.
"Auto time scaling"
defines the factor the horizontal scale is multipied
with.
"Auto time" shows the resulting holdof
f time: Auto time = Auto time
scaling * Horizontal scale.
"Off"
No holdoff
Remote command:
TRIGger<m>:
HOLDoff:
MODE on page 1420
TRIGger<m>:HOLDoff:TIME on page 1421
TRIGger<m>:HOLDoff:EVENts on page 1422
TRIGger<m>:HOLDoff:MAX on page 1423
TRIGger<m>:HOLDoff:MIN on page 1422
TRIGger<m>:HOLDoff:AUTotime? on page 1423
TRIGger<m>:HOLDoff:SCALing on page 1423
6.5 Conditioning
The rejection of noise by setting a hysteresis avoids unwanted trigger events caused
by noise oscillation around the trigger level.
You can select the hysteresis mode and value for each channel separately, or couple
the trigger levels and set the same hysteresis for channels. The hysteresis of the exter-
nal trigger input is an independent value, and it is not af
fected by level coupling.