Remote control commands
R&S
®
RTO6
1421User Manual 1801.6687.02 ─ 05
Suffix: 
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Parameters:
<Mode> TIME | 
EVENts | RANDom | AUTO | OFF
TIME
Defines the holdoff directly as a time period. The next trigger
occurs only after the "Holdof
f time" has passed (defined using
TRIGger<m>:HOLDoff:TIME).
EVENts
Defines the holdoff as a number of trigger events. The next trig-
ger occurs only when this number of events is reached. The
number of triggers to be skipped is defined using TRIGger<m>:
HOLDoff:EVENts.
RANDom
Defines the holdoff as a random time limited by TRIGger<m>:
HOLDoff:MIN on page 1422 and TRIGger<m>:HOLDoff:MAX
on page 1423. For each acquisition cycle, the instrument selects
a new random holdoff time from the specified range.
AUTO
The holdoff time is calculated automatically based on the current
horizontal scale.
OFF
No holdoff
*RST:  OFF
Usage:  Asynchronous command
TRIGger<m>:HOLDoff:TIME <T
ime>
Defines the holdof
f time period. The next trigger occurs only after this time has passed.
The setting is relevant if the holdoff mode is set to TIME.
See also:
●
TRIGger<m>:HOLDoff:MODE
Suffix: 
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<T
ime> Range:  100E-9  to  10
Increment:  200E-6
*RST:  1E-3
Default unit: s
Example: 
TRIGger:HOLDoff:MODE TIME
TRIGger:
HOLDoff:TIME 1ms
The holdoff time is set to 1 ms.
Trigger