Remote control commands
R&S
®
RTO6
1320User Manual 1801.6687.02 ─ 05
Parameters:
<State> ON | OFF
*RST: ON
Usage: Asynchronous command
DISPlay:PERSistence:INFinite <State>
If persistence is enabled (DISPlay:PERSistence[:
STATe]), each new data point in
the diagram area remains on the screen infinitely until this command is set to "OFF".
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
DISPlay:PERSistence:TIME <Time>
If persistence is enabled (DISPlay:PERSistence[:STATe]), each new data point in
the diagram area remains on the screen for the duration defined here.
Parameters:
<Time> Range: 0.05 to 50
Increment: 0.05
*RST: 0.05
Default unit: s
Usage: Asynchronous command
DISPlay:PERSistence:RESet
Resets the display
, removing persistent values.
Usage: Event
Asynchronous command
24.7.3.2 Signal
DISPlay:INTensity
<Intensity>
This value determines the strength of the waveform line in the diagram. Enter a per-
centage between 0 (not visible) and 100% (very strong).
The exact mapping of the cumulative value occurences according to the assigned color
table is guaranteed only if the intensity is set to 50% (default). All other intensity values
falsify the mapping but may improve the visibility of the signal.