Measurements
R&S
®
RTO6
347User Manual 1801.6687.02 ─ 05
The measurement parameters are common for all histograms measurements, you find
them directly on the "Setup"
tab.
Source
Selects the histogram on which the measurement is based. Before, you have to define
a histogram, see Chapter 8.2.8.3, "Creating histograms", on page 344.
Remote command:
MEASurement<m>:HISTogram:SELect on page 1528
Probability domain marker, Reference
Defines the marker reference in the probability domain.
"Peak"
The y-value with the maximum sample value in the histogram
"Upper Peak"
The y-value at the maximum sample value in the upper half of the
histogram
"Lower Peak"
The y-value at the maximum sample value in the lower half of the his-
togram
"Maximum"
The highest y-value with a probability > 0
"Minimum"
The lowest y-value with a probability > 0
"Median"
The y-value for which half the samples lie above, the other half below
in the histogram.
"Mean"
The weighted arithmetic average of the histogram
Remote command:
MEASurement<m>:
HISTogram:
PROBability:TYPE on page 1529
Probability domain marker, Delta
Defines a range around the marker.
Remote command:
MEASurement<m>:
HISTogram:PROBability:LIMit on page 1529
8.2.9 Jitter measurements
Jitter measurements are availabe if one of the following options is installed:
●
R&S RTO6-K12
Automatic measurements