Remote control commands
R&S
®
RTO6
1393User Manual 1801.6687.02 ─ 05
Parameters:
<Slope> POSitive | NEGative | EITHer
See Chapter 24.4.3, "Slope parameter", on page
1285.
*RST
: POSitive
Usage: Asynchronous command
24.9.3 Glitch trigger
TRIGger<m>:GLITch:POLarity
.......................................................................................1393
TRIGger<m>:GLITch:RANGe........................................................................................ 1393
TRIGger<m>:GLITch:WIDTh......................................................................................... 1394
TRIGger<m>:GLITch:POLarity <Polarity>
Defines the polarity of a pulse, that is the direction of the first pulse slope.
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Polarity> POSitive | NEGative | EITHer
See Chapter 24.4.4, "Polarity parameter", on page 1285.
*RST
: POSitive
Usage: Asynchronous command
TRIGger<m>:GLITch:RANGe <RangeMode>
Selects which glitches are identified: shorter or longer than the width specified using
TRIGger<m>:GLITch:
WIDTh.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<RangeMode> SHORter | LONGer
SHORter
Glitches shorter than the specified width are identified.
LONGer
Glitches longer than the specified width are identified.
*RST: SHORter
Usage: Asynchronous command
Trigger