Protocol analysis
R&S
®
RTO6
725User Manual 1801.6687.02 ─ 05
ESC Error
Searches for an escape error
.
Remote command:
SEARch:
TRIGger:SWIRe:ERRor:ESC on page 2012
13.14.5.2 SpaceWire search results
To get search results, "Enable" the search. You can minimize, shift or close the search
dialog.
The search results are listed in the search result table and marked in the waveform by
blue lines.
For an introduction to search results, see:
●
Chapter 11.1.2, "Search results", on page 420
●
Chapter 11.4.2, "Display zoom settings", on page 440
Remote commands:
●
SEARch:RESult:SWIRe:FCOunt? on page 2014
●
SEARch:RESult:SWIRe:FRAMe<m>:DATA? on page 2014
●
SEARch:RESult:SWIRe:FRAMe<m>:STARt? on page 2015
●
SEARch:RESult:SWIRe:FRAMe<m>:STATe? on page 2015
●
SEARch:RESult:SWIRe:FRAMe<m>:STOP? on page 2015
●
SEARch:RESult:SWIRe:FRAMe<m>:TYPE? on page 2016
13.14.5.3 Searching SpaceWire
Prerequisite: A serial bus is configured for the SpaceWire signal to be decoded and
analyzed.
The search for events is set up in the following way:
1. Press the [Bus] key and selet the "Shortcuts" tab.
2. Tap "Setup search".
3. If the dialog box does not contain a search entry, tap the "Add" icon to create one,
as described in "To create a user-defined search" on page 434.
4. Tap "Source" and select the serial bus that is set to SpaceWire (e.g. "SB1", unless
already selected).
5. Specify search criteria according to Chapter 13.14.5.1, "SpaceWire search setup",
on page 723.
6. To acquire a waveform, press [Single].
The R&S RTO6 performs a SpaceWire decode according to the thresholds and
protocol settings of the associated serial bus source.
SpaceWire (option R&S R
T
O6-K530)