Remote control commands
R&S
®
RTO6
1659User Manual 1801.6687.02 ─ 05
24.16.7 Long term measurement results and measurement histogram
export to file
EXPort:MEASurement:SELect.......................................................................................
1659
EXPort:MEASurement:TYPE
.........................................................................................1659
EXPort:MEASurement:NAME........................................................................................1659
EXPort:MEASurement:SAVE.........................................................................................1660
EXPort:MEASurement:DATA?....................................................................................... 1660
EXPort:MEASurement:SELect <SelcMeas>
Selects the measurement for export of long term or measurement histogram data.
Parameters:
<SelcMeas> MEAS1 | MEAS2 | MEAS3 | MEAS4 | MEAS5 | MEAS6 |
MEAS7 | MEAS8
*RST: MEAS1
Example: See "Exporting long-term measurement data to file"
on page
1264
Usage: Asynchronous command
EXPort:MEASurement:TYPE <ExportT
ype>
Y
ou can export the result data of the long term measurement, or the measurement his-
togram, or the track data.
To export the measurement histogram, it must be enabled using MEASurement<m>:
STATistics:HISTogram.
To export the long term results, the long term measurement must be enabled using
MEASurement<m>:LTMeas[:STATe].
To export a track, the track must be enabled before. Track measurements require an
option, see "Enable (Track)" on page 361.
Parameters:
<ExportType> LONGterm | HISTogram | TRACk
LONGTERM = LONGterm, HISTOGRAM = HISTogram
*RST: HISTogram
Example: See "Exporting long-term measurement data to file"
on page 1264
Usage: Asynchronous command
EXPort:MEASurement:NAME <Path>
Sets the file name and path to save the long term or measurement histogram data to.
Data management