Triggers
R&S
®
RTO6
203User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:
WINDow:
WIDTh on page 1400
±Delta
Defines a range around the "Width" value.
Remote command:
TRIGger<m>:WINDow:DELTa on page 1400
6.3.7 Timeout
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Timeout"
The timeout trigger checks if the signal stays above or below the threshold voltage for
a specified time lapse. In other words, the trigger occurs if the trigger source does not
have the expected transition within the specified time.
Trigger level
Sets the voltage level for the trigger condition. You can also drag the trigger level
marker on the display (T
A or TB on the right edge of the display). The range of the trig-
ger level is limited in a way so that always a hysteresis for stable trigger conditions is
available.
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1391
Range
Selects the relation of the signal level to the trigger level:
"Stays high"
The signal level stays above the trigger level.
"Stays low"
The signal level stays below the trigger level.
"High or low"
The signal level stays above or below the trigger level.
Remote command:
TRIGger<m>:
TIMeout:
RANGe on page 1400
Trigger types