EasyManua.ls Logo

R&S RTO6 - 24.9.22 External trigger input

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1434User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
24.9.22 External trigger input
External trigger signals, which are connected to the external trigger input, can be trig-
gered with an edge trigger
.
TRIGger<m>:ANEDge:COUPling <Coupling>
Sets the coupling for the external trigger signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Parameters:
<Coupling>
DC | DCLimit | AC
DC
Direct connection with 50 Ω termination, passes both DC and
AC components of the trigger signal.
DCLimit
Direct connection with 1 MΩ termination, passes both DC and
AC components of the trigger signal.
AC
Connection through DC capacitor, removes DC and very low-fre-
quency components.
Usage: Asynchronous command
TRIGger<m>:ANEDge:FILTer <Filter>
Sets a filter for the external trigger signal to reject high or low frequencies.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Filter> OFF |
LFReject | RFReject
OFF
The trigger signal is not filtered.
LFReject
Frequencies higher than the "Cut-of
f" frequency are rejected,
lower frequencies pass the filter.
You can adjust the "Cut-off" frequency using the TRIGger<m>:
ANEDge:CUToff:LOWPass command, the default is 50 kHz.
Trigger

Table of Contents