Remote control commands
R&S
®
RTO6
1338User Manual 1801.6687.02 ─ 05
DCLimit
Direct connection with 1 MΩ termination, passes both DC and
AC components.
AC
Connection with 1 MΩ termination, removes DC and very low-
frequency components.
CHANnel<m>:GND <State>
Connects the signal to the ground.
Suffix:
<m>
.
1..4
Parameters:
<State> ON | OFF
*RST: OFF
CHANnel<m>:SCALe <Scale>
Sets the vertical scale for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Scale> Scale value, given in Volts per division.
Range: Depends on attenuation factors, coupling, and
instrument model, see below
.
Increment: Depends on vertical and probe settings
*RST: 0.05
Default unit: V/div
Usage: Asynchronous command
CHANnel<m>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. Use the com-
mand alternativly instead of CHANnel<m>:
SCALe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Range> V
oltage range value
Range: Dependencies see below.
Increment: Depends on vertical and probe settings
*RST: 0.5
Default unit: V/div
Acquisition and setup