Remote control commands
R&S
®
RTO6
1339User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
CHANnel<m>:POSition <Position>
Sets the vertical position of the indicated channel as a graphical value.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Position> Positive values move the waveform up, negative values move it
down.
Range: -5 to 5
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CHANnel<m>:OFFSet <Offset>
The of
fset voltage is subtracted to correct an offset-affected signal. The offset of a sig-
nal is determined and set by the autoset procedure.
See also: "Offset" on page 139.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Offset> Negative values move the waveform up, positive values move it
down.
Range: Depends on attenuation factors, input coupling, and
the of
fset compensation range of active probes.
The nominal of
fset range for 1:1 attenuation and
probe offset compensation = 0 is specified in the
data sheet.
Increment: Depends on vertical and probe settings
*RST: 0
Default unit: V
Usage: Asynchronous command
CHANnel<m>:INVert <InvertChannel>
Turns the inversion of the signal amplitude on or of
f. To invert means to reflect the volt-
age values of all signal components against the ground level. If the inverted channel is
the trigger source, the instrument triggers on the inverted signal.
Acquisition and setup