Remote control commands
R&S
®
RTO6
1340User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<InvertChannel> ON | OFF
*RST: OFF
Usage: Asynchronous command
Firmware/software: FW 3.30
CHANnel<m>:BANDwidth <BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<BandwidthLimit> FULL
FULL
Use full bandwidth.
B200 | B20
Limit to 200 MHz or 20 MHz.
*RST: FULL
Usage: Asynchronous command
CHANnel<m>:CPLing <Channel>, <State>
CHANnel<m>:CPLing
? <Channel>
Sets the vertical settings of the coupled channel to the values of the active channel.
Suffix:
<m>
.
1..4
Active channel
Parameters:
<State> OFF | ON
Parameters for setting and query:
<Channel> CHAN1 | C1 | CHAN2 | C2 | CHAN3 | C2 | CHAN4 | C4
CHAN1 = C1, CHAN2 = C2, CHAN3 = C2, CHAN4 = C4
Input channel that takes the vertical settings of CHANnel<m>.
Usage: Asynchronous command
CHANnel<m>:IMPedance <Impedance>
Sets the impedance of the channel for power calculations and measurements.