Remote control commands
R&S
®
RTO6
1346User Manual 1801.6687.02 ─ 05
PROBe<m>:SETup:OFFSet:USEautozero <AutoZeroOf
fset>
Includes the AutoZero of
fset in measurement results. The auto zero error is detected
with PROBe<m>:SETup:OFFSet:AZERo.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<AutoZeroOffset> ON | OFF
*RST: OFF
Usage: Asynchronous command
Firmware/software: Version 2.70
CHANnel<m>:EATScale <ExtAttScl>
Sets the attenuation scale for an external divider.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAttScl> LIN | LOG
*RST: LIN
Usage: Asynchronous command
CHANnel<m>:EATTenuation <ExtAtt>
Sets the attenuation of an external voltage divider that is part of the DUT before the
measuring point. The external attenuation is included in the measurement, and the
instrument shows the results that would be measured before the divider.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAtt> V
alues and unit depend on the selected scale (CHANnel<m>:
EATScale).
Range: Linear scale: 1E-3 to 1E+6, logarithmic scale: -60
dB to 120 dB
Increment: 0.01
*RST: 1