Remote control commands
R&S
®
RTO6
2465User Manual 1801.6687.02 ─ 05
PSRC:REPRate <RepetitionRate>
Sets the pulse frequency
, the repetition rate of the pulse.
Parameters:
<RepetitionRate>
Available values depend on the selected clock mode:
Locked: 5/10/20/50/100/200/500 Hz, 1/5/10/25/50/100/250 MHz
Free running: 5/10/20/50/100/200/500 Hz, 1/5/10/25/50 MHz
*RST:  1E+6
Default unit: Hz
Usage:  Asynchronous command
PSRC:DUTYcycle <DutyCycle>
Sets the duty cycle of the pulse, which is the ratio of the positive pulse width to the
period of the signal.
Parameters:
<DutyCycle> For repetition rates > 5 MHz, the value is fixed at 50 %.
Range:  10  to  90
Increment:  10
*RST:  50
Default unit: %
Usage:  Asynchronous command
PSRC:CLOCkmode <ClockMode>
Sets the dependency of the pulse clock on the instrument's reference clock.
Parameters:
<ClockMode> LOCKed | FREerunning
LOCKed
The pulse source is locked to the reference clock of the instru-
ment.
FREerunning
The clock of the pulse source is independent. Deviations of the
system do not af
fect the pulse clock, and deterministic condi-
tions are avoided.
*RST
:  FREerunning
Usage:  Asynchronous command
PSRC:RST
Resets the pulse source to the default state.
Usage:  Event
Asynchronous command
Pulse Source and TDR/TDT Analysis (options R&S RT
O6-B7/K130)