Remote control commands
R&S
®
RTO6
1383User Manual 1801.6687.02 ─ 05
CHANnel<m>:DIGFilter:STATe <State>
Enables the DSP filter for input channels.
Suffix:
<m>
.
1..4
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
CHANnel<m>:DIGFilter:CUToff <CutOffLP>
Sets the limit frequency of the lowpass filter for input channels.
Suffix:
<m>
.
1..4
Selects the input channel. One filter is applied to a pair of chan-
nels - one filter for channels 1 and 2 and another filter for chan-
nels 3 and 4 (only 4-channel models).
Parameters:
<CutOffLP> Range: 100E+3 to instrument bandwidth
*RST: 1E+6
Default unit: Hz
Available filter values are:
100E+3, 200E+3, 300E+3, 500E+3, 1.0E+6, 2.0E+6, 3.0E+6,
5.0E+6, 10E+6, 20E+6, 30E+6, 50E+6, 100E+6, 500E+6, 1.0E
+9, 1.5E+9, 2.0E+9, 2.5E+9, 3.0E+9, 3.5E+9, 4.0E+9, 4.5E+9,
5.0E+9, 5.5E+9, 6.0E+9
Usage: Asynchronous command
TRIGger<m>:COUPling <DigTrigCpl>
Selects the filter for the trigger channel(s). Other channels must use the same filter
, or
proceed unfiltered.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<DigT
rigCpl> OFF | RFReject
OFF
The trigger signal is not filtered.
RFReject
Frequencies higher a given limit are rejected, lower frequencies
pass the filter. The limit is set with
TRIGger<m>:RFReject.
*RST: OFF