Parameters:
<SMIDPatternTo> Numeric or string pattern, Chapter 24.4.6, "Bit pattern parame-
ter", on page 1286
. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT
:SDCN <SMDataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<SMDataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one data pattern
to be set with
TRIGger<m>:SENT:SDMN.
INRange | OORange
In range / Out of range. To define the range set the minimum
and maximum values with TRIGger<m>:SENT:SDMN and
TRIGger<m>:SENT:SDMX.
*RST: EQUal
TRIGger<m>:SENT:SDMN <SMDataPattern>
Specifies a data pattern, or sets the start value of a data pattern range.
Parameters:
<SMDataPattern> Numeric or string pattern, Chapter 24.4.6, "Bit pattern parame-
ter", on page 1286. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT:SDMX <SMDataPatternT
o>
Sets the end value of an identifier range for TRIGger<m>:SENT:SCONdition
INRange or OORange.
Parameters:
<SMDataPatternT
o> Numeric or string pattern, Chapter 24.4.6, "Bit pattern parame-
ter", on page 1286. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT:FORMerror <FormError>
T
riggers on format errors in serial messages.
A form error occurs when at least one of the transmission sequences that form a serial
message has an error
.
Protocols