Mask testing
R&S
®
RTO6
404User Manual 1801.6687.02 ─ 05
Figure 10-3: Waveform mask with vertical width = 0.5 div
Remote command:
MTESt:WFMRescale:
YWIDth on page 1593
Vertical stretch
Sets the vertical scaling to stretch the mask in y-direction. The scaling axis is the hori-
zontal line through the lowest value of the lower mask limit. Values > 100% stretch the
mask, and values < 100% compress it.
Figure 10-4: Waveform mask with vertical width = 0.5 div, vertical position = -0.5 div, vertical stretch =
110%
Remote command:
MTESt:WFMRescale:
YSTRetch on page 1594
Vertical position
Moves the mask vertically within the display.