Remote control commands
R&S
®
RTO6
2388User Manual 1801.6687.02 ─ 05
BUS<m>:DDR:DATA:SOURce <DataSource>
Selects the source for the data output signal.
Suffix:
<m>
.
1..4
Parameters:
<DataSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 |
R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
*RST: C1W1
Usage: Asynchronous command
BUS<m>:DDR:DATA:THReshold <DataThreshold>
Sets the threshold value for the digitization of the data line.
Suffix:
<m>
.
1..4
Parameters:
<DataThreshold> Range: -10 to 10
Increment: 0.01
*RST: 0.75
Default unit: V
Usage: Asynchronous command
BUS<m>:DDR:STRBe:HYSTeresis <StrobeHyst>
Sets a value for the hysteresis of the strobe.
Suffix:
<m>
.
1..4
Parameters:
<StrobeHyst> Range: 0 to 10
Increment: 0.01
*RST: 0.35
Default unit: V
Usage: Asynchronous command
BUS<m>:DDR:STRBe:SOURce <StrobeSource>
Selects the source for the strobe signal.