Remote control commands
R&S
®
RTO6
1729User Manual 1801.6687.02 ─ 05
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:
UART:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<TxThreshold>
User-defined clock threshold
Range: -15 to 15
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:UART:TECHnology <T
echnology>
Sets the threshold voltage Tx and Rx lines as defined for various signal technologies.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<T
echnology> V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
V15 | V25 | V165 | V125 | V09 | V38 | V20 | V0
1.5 V
, 2.5 V
, 1.65 V ... respectively
VM13
-1.3 V (negative value)
MANual
Manual setting of user-defined values with BUS<m>:UART:RX:
THReshold and BUS<m>:UART:TX:THReshold.
*RST: V165
BUS<m>:UART:BITRate <Bitrate>
Sets the number of transmitted bits per second.
Suffix:
<m>
.
1..4
Parameters:
<Bitrate> Range: 300 to 20000000
Increment: 1
*RST: 9600
Default unit: bps
Usage: Asynchronous command
Protocols