EasyManua.ls Logo

R&S RTO6 - Page 1728

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1728User Manual 1801.6687.02 ─ 05
Parameters:
<RxSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 |
M7 | M8 |
R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
RX and TX lines.
See Chapter 24.4.2, "Waveform parameter", on page 1284
*RST: None
Usage: Asynchronous command
BUS<m>:UART:TX:SOURce <TxSource>
Selects the input channel for the transmitter signal.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<TxSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 |
M7 | M8 |
R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
RX and TX lines.
See Chapter 24.4.2, "Waveform parameter", on page 1284
*RST: C1W1
Usage: Asynchronous command
BUS<m>:UART:RX:THReshold <RxThreshold>
Sets a user-defined threshold value for the Rx line.
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:
UART:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<RxThreshold> User-defined clock threshold
Range: -15 to 15
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:UART:TX:THReshold <TxThreshold>
Sets a user-defined threshold value for the Tx line.