Remote control commands
R&S
®
RTO6
1804User Manual 1801.6687.02 ─ 05
TRIGger<m>:CAN:FDATa:BRS <BRSBit>
Sets the bit rate switch bit.
Parameters:
<BRSBit> ONE | ZERO | DC
ONE: the bit rate switches from the bit rate of the arbitration
phase to the faster data rate.
*RST: ONE
TRIGger<m>:CAN:FDATa:ESI <ESIBit>
Sets the error state indicator bit.
Parameters:
<ESIBit> ONE | ZERO | DC
DC: don't care, bit is nor relevant
*RST: DC
TRIGger<m>:CAN:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | L
THan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with
TRIGger<m>:CAN:DMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:CAN:DMIN and TRIGger<m>:
CAN:DMAX.
*RST: EQUal
TRIGger<m>:CAN:DMIN <DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).