2003User Manual 1801.6687.02 ─ 05
TRIGger<m>:SWIRe:TIME:MAX....................................................................................
2005
TRIGger<m>:
SWIRe:TIME:MIN.....................................................................................2005
TRIGger<m>:SWIRe:TYPE...........................................................................................2005
TRIGger<m>:SWIRe:CTYPe <ControlType>
Triggers on a specific control type character.
Suffix:
<m>
.
1..3
Parameters:
<ControlType> ANY | FCT | EOP | EEP
FCT
Flow Control T
oken
EOP
Normal End of Packet
EEP
Error End of Packet
*RST
: ANY
TRIGger<m>:SWIRe:DATA:CONDition <DataOperator>
Set the condition for the data value. You can define an exact value or a value range.
Suffix:
<m>
.
1..3
Parameters:
<DataOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | L
THan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:
SWIRe:DATA:
MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:SWIRe:DATA:MIN and
TRIGger<m>:SWIRe:DATA:MAX.
*RST: EQUal
TRIGger<m>:SWIRe:DATA:MAX <DataPatternMax>
Sets the the end value of a data for the data pattern range if TRIGger<m>:SWIRe:
DATA:CONDition is set to
INRange or OORange.
Suffix:
<m>
.
1..3